5626/2628 platinum resistance thermometer from Fluke Calibration meets all ITS-90 requirements for resistance ratios & has an accuracy of ± 0.006 °C at 0 °C.

Product Overview

If you’re using block calibrators, furnaces, or temperature points above normal PRT (platinum resistance thermometer) temperatures (420 °C), then these two PRTs are for you. The Fluke 5626 is nominally 100 Ω and the 5628 (formerly Hart Scientific) is nominally 25.5 Ω. Both instruments have a temperature range of –200 °C to 661 °C. They make great working or check standards for calibration work up to the aluminum point.

Using a regular PRT at temperatures above 500 °C exposes the platinum to contamination. If the PRT is used as a reference or calibration standard, contamination is a major problem. SPRTs (standard platinum resistance thermometers), which are more expensive and delicate, can handle the higher temperatures, but with greater risk to the instrument due to shock, contamination, or mishandling. The 5626 and 5628 are designed to reduce the contamination risk through the use of internal protection while not impairing performance.

In addition to the right measurement performance and durability, a PRT for secondary applications should be priced affordably. Fluke Calibration’s new PRTs are inexpensive and come with an accredited calibration. The calibration comes complete with ITS-90 constants and a resistance-versus-temperature table.

Check the temperature range, check the stability, check the price! Who else gives you this much quality, performance, and value for your money? No one!

Models
  • 5626-20-X
  • 5628-20-X
  • 5626-12-X
  • 5626-15-X
  • 5628-12-X
  • 5628-15-X

“X” in a model number denotes the probe termination (e.g. A = INFO-CON for 914X, B = bare wire, D = 5-pin DIN, G = gold pins, J = banana plugs, L = mini spade lugs, M = mini banana plugs, P = INFO-CON for 1523/1524, S = spade lugs). Please specify when enquiring.

Specifications: 5626, 5628 | Secondary Standard PRTs

Specifications
Temperature Range –200 °C to 661 °C
Handle Temp. 0 °C to 80 °C
RTPW 5626: 100 Ω (± 1 Ω)
5628: 25.5 Ω (± 0.5 Ω)
Resistance Ratio W(Ga) W(302.9146K) ≥ 1.11807
α ≥ 0.003925
Calibrated Accuracy† (k=2) ± 0.006 °C at –200 °C
± 0.006 °C at 0 °C
± 0.015 °C at 420 °C
± 0.022 °C at 661 °C

Stability
5626: ± 0.003 °C
5628: ± 0.002 °C
Long-Term Drift (k=2) 5626: < 0.006 °C/100 hours at 661 °C
5628: < 0.004 °C/100 hours at 661 °C
Immersion At least 12.7 cm (5 in) recommended
Sheath Inconel™ 600
Lead Wires 4-wire Super-Flex PVC, 22 AGW
Termination Gold-plated spade lugs, or specify
Size 6.35 mm dia. x 305 mm, 381 mm, or 508 mm (0.25 x 12, 15, or 20 in) standard, custom lengths available
Calibration Accredited calibrations from Fluke Calibration
†Includes calibration and 100 hr drift

Get in touch

We're available 24/7 to assist. Just let us know what we can help you with and we'll do our best.

We'll aim to get back to you within 24 hours.