Compact split-core fluxgate current sensors for 500 A and 200 A. They clamp onto an existing conductor and combine power-analyzer accuracy with enough bandwidth for waveform instruments. Voltage output on a BNC connector works with WT series and PX8000 power analyzers and with ScopeCorders and DLM oscilloscopes.
Models & Specifications
Common Specifications
| Specification |
Value |
| Output resistance |
50 Ω ±10 Ω |
| Supply |
±12 V |
| Connector |
BNC |
| Main unit dimensions |
110 × 62 × 25 mm |
| Operating temperature |
Main unit −40 to 85 °C; sub unit and IV unit 5 to 40 °C |
| Phase accuracy |
±0.1° from 0.1 Hz to 1 kHz |
| Accuracy warranty |
12 months |
Model Comparison
| Specification |
CT500SA |
CT200SA |
| Rated current |
500 A AC/DC |
200 A AC/DC |
| Bandwidth (−3 dB) |
DC to 500 kHz, typical |
DC to 1 MHz, typical |
| Output |
4 mV/A, BNC |
10 mV/A, BNC |
| Delay |
130 ns typical |
78 ns typical |
| Supply current |
0.8 A maximum |
0.7 A maximum |
| Aperture |
20 mm |
20 mm |
Specifications from the official Yokogawa user’s manual IM CT500SA-01EN. Use the Download Datasheet button for the full document.