Through-hole AC/DC current sensor for the largest currents on a WT series power analyzer or PX8000. It measures 0 to 2000 A DC and 2000 A rms (3000 A peak) AC, with basic accuracy of ±(0.05% of reading + 30 µA) and bandwidth from DC to 40 kHz.
When several CT2000A sensors are powered from the WT1800E or PX8000 /PD2 option, keep the total supply current within 6 A (WT1800E, channels 1 to 6) or 4 A (PX8000, channels 1 to 4).
Models & Specifications
| Specification |
CT2000A |
| DC current |
0 to 2000 A |
| AC current |
2000 A rms (3000 A peak) |
| Bandwidth (−3 dB) |
DC to 40 kHz |
| Basic accuracy |
±(0.05% of reading + 30 µA) |
| Transformation ratio |
2000:1 |
| Output at rated current |
1 A at 2000 A |
| Continuous maximum input |
3000 A peak |
| Instantaneous maximum |
10000 A peak, 0.1 s |
| Load resistance |
0 to 1 Ω |
| Aperture |
70 mm |
| Supply |
±(15 V ±5%) |
| Output connector |
D-sub 9-pin |
| Operating temperature |
−40 to 85 °C |
| Maximum rated voltage |
1000 V rms, CAT III |
| Dimensions |
230 × 220 × 76 mm |
| Weight |
Approx. 4.2 kg |
| Maximum rated power |
35 VA |
Specifications from the official Yokogawa user’s manual IM CT2000A-01EN. Use the Download Datasheet button for the full document.