Fluke

9009 Dual-Well Dry-Well

跳至产品信息

Model

对这款产品感兴趣?

请留下您的联系方式,我们会尽快回复库存、报价及所需信息。

The Fluke Calibration 9009 dry-well offers two independently controlled temperature blocks (-15°C to 350°C) in a portable design for in-field calibrations.

Product Overview

Two blocks. One portable calibrator. Full temperature coverage.

Dual Independent Temperature Blocks

The Fluke Calibration 9009 combines two independently controlled temperature blocks into one compact, rugged system:

  • Hot block range: 50 °C to 350 °C
  • Cold block range: –15 °C to 110 °C
  • Each block is powered by a precision Hart Scientific temperature controller—not off-the-shelf components, but industry-leading technology from the world’s top temperature calibration experts.

Flexible Multi-Probe Calibration

Each block includes two wells with removable inserts, giving you unmatched flexibility:

  • Calibrate up to four probes simultaneously
  • Use two probes with an external reference thermometer (e.g., Hart 1521 LLK)
  • Perform quick zero/span checks for transmitter calibration with both blocks in use

Portable and Field-Ready

The 9009 is built for portability and field durability:

  • Housed in a rugged Pelican™ case—airtight, watertight, and impact-resistant
  • Weighs only 10 lbs (4.5 kg), compact and easy to carry
  • Internal storage holds the power cord and four extra inserts
  • Inserts available for sensor sizes from 1/16” (1.6 mm) to 7/16” (11.1 mm)

This system is truly ready to go wherever your work takes you.

Reliable Accuracy and Stability

Performance you expect from a Fluke/Hart Scientific source:

Block Display Accuracy Stability
Cold ±0.2 °C ±0.05 °C
Hot ±0.6 °C ±0.05 °C

NIST-traceable calibration certificate included for both blocks.

PC Connectivity and Automation

  • RS-232 interface included for temperature monitoring and control via PC
  • Fully compatible with MET/TEMP II software for automated calibration of RTDs, thermocouples, and thermistors

The Complete Field Calibration Package

With the Fluke 9009 Dual-Well Dry-Well, you get:

  • Two independent blocks
  • Full temperature coverage from –15 °C to 350 °C
  • True field portability in a compact, durable case
  • Flexibility for multiple sensors and calibration modes
  • Built-in automation and PC integration
Models
  • 9009-B-156
  • 9009-Y-156

Specifications: Fluke Calibration 9009 Dual-Well Dry-Well

General Specifications
Hot block Cold block
Range 50°C to 350°C (122°F to 662°F) –15°C to 110°C (5°F to 230°F)
(–8°C [18°F] with hot block at 350°C [662°F])
Accuracy ±0.6°C ±0.2°C
Stability ±0.05°C
Well-to-well uniformity ±0.1°C
Display resolution 0.1°
Heating times 30 minutes from 25°C to 350°C 15 minutes from 25°C to 110°C
Cooling times 40 minutes from 350°C to 100°C 16 minutes from 25°C to –15°C
Stabilization times 8 minutes
Well depth 4 in (102 mm)
Removable inserts Two 6.4 mm (1/4 in) and two 4.8 mm (3/16 in) inserts included; see ordering information for other available inserts
Computer interface RS-232 included
Power 115 V AC (±10 %), 3 A, or
230 V AC (±10 %), 2 A, specify, 50/60 Hz, 280 W
Size (H x W x D) 178 x 267 x 248 mm (7 x 10.5 x 9.75 in)
Weight 4.5 kg (10 lb)
NIST-traceable calibration Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C Data at –8°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 110°C

联系我们

我们随时为您提供协助。请告诉我们您的需求,我们会尽力帮助。