Fluke

5627A Precision Industrial PRT

跳至产品信息
1 / 4

Model

对这款产品感兴趣?

请留下您的联系方式,我们会尽快回复库存、报价及所需信息。

Fluke Calibration's 5627A temperature RTD probe is NVLAP-accredited, has a temperature range up to 420 °C & an accuracy as good as ± 0.025 °C.

Product Overview

One of the best features of this RTD sensor is that it conforms to the standard 385 curve, letting you use your DIN/IEC RTD meters fully. Why use a precision thermometer or RTD temperature probe that’s less accurate than your meter?

The 5627A is manufactured using a coil suspension element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 19 mm (3/4-inch) for flexibility and durability. (Bend, if any, should be specified at time of order.)

Six- and nine-inch 5627A precision thermometers are calibrated at –196 °C, –38 °C, 0 °C, 200 °C, and 300 °C. For 12-inch versions of these precision thermometers, the point at 300 °C is replaced by a calibration point at 420 °C.

Each probe is individually calibrated and includes an accredited report of calibration from the manufacturer.

This RTD temperature probe is an excellent value for a precision thermometer. It has the price-to-accuracy and price-to-durability ratios you should demand in all of the precision thermometers you buy!

Models
  • 5627A-6-X
  • 5627A-9-X
  • 5627A-12-X

“X” in a model number denotes the probe termination (e.g. A = INFO-CON for 914X, B = bare wire, D = 5-pin DIN, G = gold pins, J = banana plugs, L = mini spade lugs, M = mini banana plugs, P = INFO-CON for 1523/1524, S = spade lugs). Please specify when enquiring.

Specifications: 5627A | Precision Thermometer, RTD Temperature Probe

Specifications
Resistance Nominal 100 Ω
Temperature Coefficient 0.00385 Ω/Ω/ °C nominal
Temperature Range −200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C)
Drift Rate (k=2) ± 0.04 °C at 0 °C after 100 hours at 420 °C
Sheath Material 316 Stainless Steel
Leads PTFE-insulated, nickel-plated stranded copper, 22 AWG
Termination Specify. See Ordering Information.
Time Constant Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
Bending Radius Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Fluke lab requires 20 cm [8 in] of unbent sheath to re-calibrate.)
Calibration Includes manufacturer's accredited calibration and table with R vs. T values in 1 °C increments from -196 °C to 420°C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included.
Immersion At least 100 mm (4 in) recommended
Calibrated Accuracy† (k=2) ± 0.026 °C at -196 °C
± 0.046 °C at 0 °C
± 0.077 °C at 200 °C
± 0.124 °C at 420 °C
Size 5627A-12: 305 mm x 6.35 mm (12 in x 1/4 in)
5627A-9: 229 mm x 4.7 mm (9 in x 3/16 in)
5627A-6: 152 mm x 4.7 mm (6 in x 3/16 in)
†Includes calibration uncertainty and 100 hr drift.

联系我们

我们随时为您提供协助。请告诉我们您的需求,我们会尽力帮助。

我们会尽量在 24 小时内回复您。