Need support?
Contact us by email directly
Your cart is empty :(
Continue shoppingYour cart
Estimated total:
HK$0.00
Send us your details and we’ll get back to you with availability, pricing, and any additional information you need.
The 40BD is an IEC 61094 WS3P ¼″ prepolarized pressure microphone with rear-venting. It is a high-precision condenser microphone made according to IEC 61094-4 requirements. Its low sensitivity makes it ideal for sound measurements at high sound pressure levels of up to 174 dB. Its small size reduces the effects of diffraction and reflections around the microphone, resulting in a frequency range that extends up to 70 kHz. NOTE: Optimized for use without protection grid—remove grid before measuring.
40BD is individually factory calibrated and delivered with a calibration chart stating its specific open-circuit sensitivity and pressure frequency response.
NOTE: At low frequencies, protection grids are acoustically transparent; however, around 20 kHz and above, protection grids can act as resonators and influence the frequency response. So for higher frequency measurements, ¼ ″ and ⅛ ″ microphones have been optimized for use without the protection grid. Please remove the protection grid when measuring.
| Polarization/Connection | 0 V / CCP |
| Frequency range (±1 dB) | 10 Hz to 25 kHz |
| Frequency range (±2 dB) | 5 Hz to 70 kHz |
| Dynamic range lower limit (microphone thermal noise) | 34 dB(A) |
| Dynamic range lower limit with GRAS preamplifier | 44 dB(A) |
| Dynamic range upper limit with GRAS preamplifier @ +28 V / ±14 V power supply | 166 dB |
| Dynamic range upper limit with GRAS preamplifier @ +120 V / ±60 V power supply | 174 dB |
| Dynamic range upper limit with GRAS preamplifier | 166 dB |
| Open-circuit sensitivity @ 250 Hz (±3 dB) | 1.6 mV/Pa |
| Open-circuit sensitivity @ 250 Hz (±3 dB) | -56 dB re 1V/Pa |
| Resonance frequency | 50 kHz |
| Microphone cartridge capacitance, typ. | 7 pF |
| Microphone venting | Rear |
| Temperature range, operation | -40 to 120 °C / -40 to 248 °F |
| IEC 61094-4 Designation | WS3P |
| Temperature range, storage | -40 to 85 °C / -40 to 185 °F |
| Temperature coefficient @250 Hz | -0.01 dB/°C / -0.006 dB/°F |
| Humidity range non condensing | 0 to 100 % RH |
| Static pressure coefficient @250 Hz | -0.008 dB/kPa |
| Humidity coefficient @250 Hz | -0.001 dB/% RH |
| Influence of axial vibration @1 m/s² | 55 dB re 20 µPa |
| Weight | 1.5 g / 0.053 oz |
Specifications from the official GRAS product sheet. Please refer to the attached datasheet for full details.