{"product_id":"1595a-super-thermometer","title":"1595A Super-Thermometer","description":"\u003cp\u003eThe Fluke Calibration 1595A Super-Thermometer delivers unprecedented accuracy, low noise, and fast measurements for high-precision thermometry applications.\u003c\/p\u003e\u003ch2\u003eProduct Overview\u003c\/h2\u003e\u003ch2\u003eResistance ratio or absolute resistance measurements: the choice is yours\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eWhen your goal is to achieve the highest measurement accuracy possible, you will mostly likely make a resistance ratio measurement (Rx\/Rs). The measurement accuracy of the 1595A over typical thermometry ratios (0.25 to 4.0) is as good as or better than 0.2 ppm. And for ratios near unity (0.95 to 1.05), the resistance ratio accuracy of the 1595A is as good as 0.06 ppm. You get the linearity of a traditional resistance bridge with an instrument that is easier to use and a much better value.\u003c\/p\u003e\u003cp\u003eExample 1: Using a 25 Ω SPRT with a 25 Ω external reference resistor, the resistance ratio uncertainty of the 1595A at the triple-point of water is only 0.06 ppm (Rx\/Rs ≈ 1). This becomes significant when considering that the uncertainty of the RTPW measurement is propagated throughout all ITS-90 temperature measurements.\u003c\/p\u003e\u003cp\u003eExample 2: The Direct Comparison Method measures the probe under test directly against a calibrated reference SPRT by defining the SPRT as the reference resistor (Rs). Since the ratio Rx\/Rs ≈ 1 over the full temperature calibration range, the uncertainty contribution of the 1595A to the overall measurement uncertainty is no greater than 0.06 ppm. Considering all other influence effects, total measurement uncertainty can be achieved in the sub-mK range.\u003c\/p\u003e\u003cp\u003eAlthough a ratio measurement can produce the most accurate results, it may not always be the most convenient measurement method. Using the temperature-controlled internal reference resistors, the Super-Thermometer can convert an absolute resistance measurement to a temperature unit and display the result in degrees Centigrade, Fahrenheit, or Kelvin, or in Ohms. Each Super-Thermometer includes internal reference resistors of 1 Ω, 10 Ω, 25 Ω, 100 Ω and 10 k Ω that support a variety of PRTs, RTDs and thermistors.\u003c\/p\u003e\u003cp\u003eHow well a thermometry bridge can measure absolute resistance depends on the bridge’s ratio accuracy, as well as the long-term stability and calibration accuracy of its internal reference resistors. To ensure stability and eliminate errors, the internal reference resistors in the Super-Thermometer are enclosed in an oven that is temperature controlled at 30 °C and stable to within approximately 10 millikelvin. In fact, these precision resistors are controlled so well that within a 24-hour period, their resistance will not change more than 0.25 ppm (equivalent to 0.00006 °C). The one-year absolute accuracy of the Super-Thermometer is 4 ppm (equivalent to 0.001 °C).\u003c\/p\u003e\u003cp\u003eMeasurement current accuracy is important when evaluating measurement uncertainty due to self heating of the thermometer. The accuracy of the current source in the Super-Thermometer is 0.2% when measuring 25 Ω or 100 Ω PRTs with a typical sensing current level.\u003c\/p\u003e\u003ch2\u003eLow measurement noise\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eMeasurement noise is caused by electrical noise and other random errors that can negatively influence measurement accuracy. New patent-pending design innovations incorporated into the Super-Thermometer reduce measurement noise and interference to levels that are unprecedented for a digital thermometry bridge. Two identical amplifier\/ADC blocks measure Rx and Rs simultaneously. Reversing the sensing current and then averaging the two measurements reduce errors caused by thermoelectric EMFs, current source instability, and electrical noise. To further reduce measurement noise, two parallel ADCs are used in each amplifier\/ADC block. Passive and active filters are also utilized to reject the majority of electrical noise and interference. In a typical temperature calibration application, the Super-Thermometer offers uncertainty due to measurement noise as low as 0.00002 °C. Under similar measurement parameters, the noise level of the Super-Thermometer can be as low as a traditional, much more expensive, resistance bridge.\u003c\/p\u003e\u003ch2\u003eUnprecedented measurement speed\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eSuper-Thermometers offer speeds as fast as one second per measurement (full accuracy with a two second sample rate). You can complete your tests in less time, track temperature changes more accurately, and even evaluate thermal response in sensors. When measuring fixed point temperatures that require the highest levels of accuracy and lowest levels of noise, you have complete control to change the measurement speed to meet your requirements.\u003c\/p\u003e\u003ch2\u003eVerify ratio accuracy in-house with Ratio Self-Calibration\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eThe 1595A features patent-pending “Ratio Self-Calibration” that enables you to test the accuracy or calibrate the resistance ratio linearity of the Super-Thermometer’s measurement circuit – regularly, easily, and without requiring any external devices or special training. The Super-Thermometer performs the Ratio Self-Calibration by automatically combining and switching between the internal temperature-controlled reference resistors to create a resistance voltage divider network. Replacing RX and RS with the resistance voltage divider allows the Super-Thermometer to complete a series of eight resistance ratio tests, from which linearity errors are calculated and accuracy relative to specifications can be verified. See figures 1 and 2 for examples of the resulting report.\u003c\/p\u003e\u003cp\u003eNow you can verify accuracy and performance automatically in about thirty minutes, at the touch of a button. Although rare, measurement circuit defects are detected and reported automatically. Through a password-protected procedure, you can choose whether to align the Super-Thermometer by applying the offsets generated by the calibration procedure. No other temperature measurement instrument on the market more fully empowers the metrologist to determine and improve instrument performance.\u003c\/p\u003e\u003cp\u003e\u003cstrong\u003eRatio Self-Calibration features at a glance\u003c\/strong\u003e\u003c\/p\u003e\u003cul\u003e\n\u003cli\u003eAutomatically test or calibrate the Super-Thermometers’ resistance ratio linearity without expensive external equipment\u003c\/li\u003e\n\u003cli\u003eDetect possible equipment failures before they affect your measurements\u003c\/li\u003e\n\u003cli\u003eNo special operator training required\u003c\/li\u003e\n\u003cli\u003eComplete the test in approximately 30 minutes\u003c\/li\u003e\n\u003cli\u003eExtend interval between calibration cycles\u003c\/li\u003e\n\u003c\/ul\u003e\u003ch2\u003eCalibrate internal reference resistors, quickly and easily\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eYou can also calibrate the Super-Thermometer’s internal reference resistors using the resistance calibration function. This process requires an external standard resistor that is approximately equal in value to the internal reference resistor to be calibrated.\u003c\/p\u003e\u003cp\u003eThe Super-Thermometer assists you with the calibration through an easy-to-use setup utility. Just specify which internal resistor is to be calibrated, the calibrated resistance value of the standard resistor, and the channel to which the standard resistor is connected. The Super-Thermometer configures the sensing current, disables filtering, and sets the timing parameters to the “precision” settings for consistent results.\u003c\/p\u003e\u003cp\u003eOnce the calibration is complete, the Super-Thermometer reports the results to the display and gives you the option to write the results to a USB memory device. You can also choose whether or not to adjust the calibration parameters of the internal reference resistor by pressing the password-protected Adjust Resistor function key.\u003c\/p\u003e\u003ch2\u003eCalculate the effects of self-heating, quickly and easily\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eWhen current is passed through a PRT sensor, power is dissipated by the sensing element, causing the sensor to self-heat. That introduces a small temperature error into the measurement. The error can be estimated by measuring the resistance of the sensor at a given temperature with two different sensing current power levels, nominal current and “double-power” (nominal current ). Through linear extrapolation to “zero-power,” the resistance value of the sensor can then be estimated as if no current were applied to the sensing element. The temperature error due to self-heating can then be effectively eliminated from the measurement.\u003c\/p\u003e\u003cp\u003eComputing a zero-power measurement manually can be time consuming and subject to calculation errors. The Super-Thermometer’s Zero-Power Measurement function sets current levels and collects measurement data automatically, calculating the zero-power measurement for you. User-adjustable settings give you full control of the process by allowing you to set parameters such as settle time, measure time, and record result.\u003c\/p\u003e\u003ch2\u003eView key measurement data in graphic or table formats, on multiple channels -- simultaneously\u003c\/h2\u003e\u003cp\u003e\u003cbr\u003eSuppose you want to see the results of more than one channel simultaneously. The Super-Thermometer includes two measurement viewing modes—graph and table view modes. Select graph mode to graph a single channel or multiple channels simultaneously; set the duration of the graphing window; select automatic centering or enter a fixed value for the vertical center; select automatic scaling; or enter a fixed value for the vertical span. Configure the graph however you like to fit your application. In table mode, the measured value, mean and standard deviation for all channels are displayed simultaneously in a numeric table format. Simply press the function key to toggle between the graph and table views.\u003c\/p\u003e\u003cdiv\u003eModels\u003cul\u003e\u003cli\u003e1595A\u003c\/li\u003e\u003c\/ul\u003e\n\u003ch2\u003eSpecifications: 1595A Super-Thermometer\u003c\/h2\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e1595A General Specifications\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMeasurement capability\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e4-wire PRT, Thermistor, Resistance, Resistance Ratio\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput resistance range\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0 to 500 kΩ\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRatio range\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0 to 10\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccepted external RS range\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e1 Ω to 10 kΩ\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInternal RS\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e1 Ω, 10 Ω, 25 Ω, 100 Ω, and 10 kΩ\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePRT conversion types\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eITS-90, PT-100, CVD-ABC, CVD-ALPHA, Polynomial\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eThermistor conversion types\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eR(T) Polynomial, T(R) Polynomial\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDisplay units\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eratio (RX \/ RS), K, °C, °F, Ω\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDisplay resolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.1 to 0.000001\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eSample period\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e(seconds)1 1, 2, 5, and 10\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eStatistics\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eAverage, Std Dev, SE of Mean, Max, Min, Difference, Peak-Peak, Delta, N\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eFront panel channels\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eFour PRT\/Thermistor inputs (channels 2 \u0026amp; 4 can be configured for either RX or RS inputs)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRear panel channels\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eTwo dedicated reference resistor inputs (RS)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput terminals\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eDWF Connector, Tellurium Copper\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eData logging to internal memory\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e80,000 individual time- and date-stamped readings (~6 MB)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInternal real-time clock\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eYes\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eComputer communications\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eRS-232, USB, IEEE-488, Ethernet\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDisplay type\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eFull VGA, LCD\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eUser interface languages\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eEnglish, French, Spanish, German, Russian, Chinese, Japanese\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003eResistance Ratio Accuracy, 95% Confidence Level, 1-yr\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e.\u003c\/td\u003e\n\u003ctd\u003e1595A\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRatio: 0.95 to 1.05\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.06 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRatio: 0.5 to 0.95, 1.05 to 2.0\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.16 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRatio: 0.25 to 0.5, 2.0 to 4.0\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.2 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRatio: 0.0 to 0.25\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.05 ppm of 1.0\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRatio: 4.0 to 10.0\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.5 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e1595A Absolute Resistance Accuracy, 95% Confidence Level, 1-yr\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e\u003cstrong\u003e(RS, current)\u003c\/strong\u003e\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0 Ω to 1.2 Ω (1 Ω, 10 mA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eGreater of 40 ppm or 0.000012 Ω\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0 Ω to 12 Ω (10 Ω, 3 mA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eGreater of 10 ppm or 0.000024 Ω\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0 Ω to 120 Ω (25 Ω, 1 mA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eGreater of 5 ppm or 0.000024 Ω\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0 Ω to 400 Ω (100 Ω, 1 mA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eGreater of 4 ppm or 0.00008 Ω\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0 kΩ to 10 kΩ (10 kΩ, 10 µA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eGreater of 5 ppm or 0.000012 Ω\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e10 k to 40 kΩ (10 kΩ, 10 µA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e8 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e40 kΩ to 100 kΩ (10 kΩ, 2 µA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e20 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e100 kΩ to 500 kΩ (10 kΩ, 1 µA)\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e80 ppm\u003csup\u003e3\u003c\/sup\u003e\n\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"3\"\u003e1595A Internal Resistor Stability\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e.\u003c\/td\u003e\n\u003ctd\u003e24 hours\u003c\/td\u003e\n\u003ctd\u003e30 days\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e1 Ω\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e5 ppm\u003c\/td\u003e\n\u003ctd\u003e10 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e10 Ω\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.5 ppm\u003c\/td\u003e\n\u003ctd\u003e2 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e25 Ω\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.25 ppm\u003c\/td\u003e\n\u003ctd\u003e1 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e100 Ω\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.2 ppm\u003c\/td\u003e\n\u003ctd\u003e1 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e10 kΩ\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.25 ppm\u003c\/td\u003e\n\u003ctd\u003e1 ppm\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e1595A Temperature Measurement Noise - Typical Performance (standard error of the mean, °C)\u003csup\u003e2\u003c\/sup\u003e\n\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e25Ω SPRT at 0°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.00002\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e25Ω SPRT at 420°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.00006\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e100Ω PRT at 0°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.00001\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e100Ω PRT at 420°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.00003\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eThermistor at 25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.000003\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e1595A Measurement Current Accuracy (self-heating)\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0.001 mA to 0.005 mA\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.00005 mA\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0.005 mA to 0.02 mA\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e1 %\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0.02 mA to 0.2 mA\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.5 %\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e0.2 mA to 2 mA\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.2 %\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003e2 mA to 20 mA\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0.5 %\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e1595A Mechanical Specifications\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e7.5 kg (16.5 lbs)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eExternal Width x Height x Length\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e432 mm x 153 mm x 432 mm (17 x 6 x 17 inches)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\u003ctd colspan=\"2\"\u003e1595A Operating Specifications\u003c\/td\u003e\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAC power requirements\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e100V to 230V ±10%, 50\/60Hz\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eSpecified operating temperature\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e15°C to 30 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMax operating temperature\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e5°C to 40 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eStorage temperature\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0°C to 40 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eOperating relative humidity, 5°C to 30°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e10% to 70%\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eOperating relative humidity, 30°C to 40°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e10% to 50%\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eStorage relative humidity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e0% to 95%, non-condensing\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eOperating altitude\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e3000 m\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWarranty period\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e1 Year\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eReport of Calibration\u003csup\u003e3\u003c\/sup\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003eNVLAP accredited\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\u003cp\u003e\u003csup\u003e1\u003c\/sup\u003e Two second sample rate required for full accuracy \u003csup\u003e2\u003c\/sup\u003e Due to the subjective nature of measurement noise, it is not a guaranteed specification \u003csup\u003e3\u003c\/sup\u003e The standard Report of Calibration includes resistance data from 1 Ω to 100 kΩ. A Report of Calibration may be ordered if resistance data from 100 kΩ to 500 kΩ is desired (see model numbers 1994 and 1995 in the ordering information table).\u003c\/p\u003e\n\u003cdiv role=\"tabpanel\" aria-labelledby=\"models-label\" class=\"tab-pane\"\u003e\u003c\/div\u003e\n\u003c\/div\u003e","brand":"Fluke","offers":[{"title":"1595A","offer_id":54038942384422,"sku":"1595A","price":0.0,"currency_code":"HKD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0953\/9797\/0214\/files\/0cc3c644-b51a-47c5-b70a-b2a30124b6f7_original__size.jpg?v=1791380880","url":"https:\/\/nplab.shop\/products\/1595a-super-thermometer","provider":"NP LAB","version":"1.0","type":"link"}